APPENDIX N BULK MILK TANKER SCREENING TEST FORM
Charm SL (raw bovine and caprine) and SL-6 (raw bovine)
Beta-Lactam Tests
GENERAL REQUIREMENTS
- __________
See Appendix N General Requirements form items 1-8 & 15
SAMPLES
- __________
See Appendix N General Requirements (GR) form item 9
APPARATUS & REAGENTS
- __________
Equipment
- __________ Strip Incubator, 56±1 0 C SL Beta-lactam test;
45±2 0 C SL–6 Beta-lactam Test
- __________
Level. Temperature checked daily (day of use), records maintained
- __________
thermometer, for each incubator (App N GR #3)
- __________
Lid closed (spring ajar), when not running tests
- __________
Incubator Temperature: ______________
- __________
ROSA Reader or equivalent with print out or download of data. ROSA Reader V1.07
or higher for SL–6, or equivalent capable of reading
3 line test, Manual Available Serial Number: ______________
- __________
Calibrators – 2 for each assay– 2 lines for SLBL
and/or 3 lines for SL–6.
Range(s)
Two Line:
Low:_________________________
High_________________________
Three Line:
Low:_________________________
High:________________________
- __________
Calibrator serial numbers match ROSA reader SN
- __________
Do not proceed if out of range. Manufacturer should be contacted for corrective actions
- __________
Printer or computer link for hardcopy download
- __________
Records maintained
- __________
Pipettor- 300 µL and disposable tips, calibrated (see App. N GR item 7)
- __________
Or s ingle use 300 µL poly-pipet with overflow bulb to accurately measure amount of sample, supplied by manufacturer (screening only)
- __________
Optional Timer – if not built into Incubator
- __________
Optional Centrifuge (Not applicable to SL–6 Beta-lactam Test)- mini or equivalent (1200–
2000 x g) for frozen controls
- __________
Reagents
- __________ Test Strips
__________Lot #: _______ Exp. Date: _______
__________Rec'd: _______ QC Date: _______by _______
- __________
Positive Control
- __________
Lyophilized or tablet 5 ppb Penicillin G for SL Beta–lactam Test
__________Lot #: _______ Exp Date: _______ ________
- __________
Lyophilized or tablet 5ppb Penicillin G/10ppb cloxacillin for SL–6 Beta-lactam Test
__________Lot # _______ Exp. Date:_______ ________
- __________
Do not proceed if out of range. Manufacturer should be contacted for corrective actions
- __________
Printer or computer link for hardcopy download
- __________
Records maintained
- __________
Negative Control
- __________
Previously negative tested raw milk (item 5b)
- __________
Reagent stability
- __________ SL–6 reagents received refrigerated.
- __________
Reagents stored at 0–4.4 0 C, maintain no longer than manufacturer's expiration date
- __________
Negative Control – raw milk tested –600 or more negative with SL Beta-lactam Test or –400 or morenegative with SL–6 Beta–lactam
__________Test Value:________ Do not proceed if out of range.
- __________
Used within 72 hours when maintained at 0–4.4 0 C
- __________
Test for suitability each time, records maintained
- __________
Or, frozen immediately (within 6 hours) and stored in a non-frost-free freezer, or in a styrofoam container in a frost free freezer, for no more than 2 months at –15 0 C or below
Lab
__________Date prep:._______ Lab Exp. Date: _______ ________
- __________
For SLBL ONLY, not applicable to SL-6 thaw, centrifuge 3 minutes, cool and use within 24 hours. Do not refreeze.
- __________
Positive Control- Manufacturer supplied, maintain no longer than manufacturer's expiration date
- __________
Reconstituted with Negative Control (raw milk), used within 48 hours when maintained at 0–4.4 0 C
- __________
Test for suitability each time prepared and/or day of use, must produce +400 or greater reading, records maintained.
__________Test Value:________ Do not proceed if out of range. ________
- __________
Or, frozen immediately (within 6 hours) and stored in a non-frost-free freezer, or in a styrofoam container in a frost-free freezer, for no more than 2 months at –15 0 C or below
__________Lab Date prep: _______ Lab Exp. Date: _______ ________
- __________
Thaw, centrifuge 3 minutes( not applicable to SL–6 Beta-lactam Test), cool and use within 24 hours. Do not refreeze.
- __________
Daily Performance and Operation Check
- __________
See App. N GR item 10)
- __________
Reader Versions 1.5 and higher, use Escape 5 reader function to enter performance monitor mode of reader- reference manual for directions
- __________
Check Calibrators, records maintained
- __________
Controls give appropriate readings in performance monitoring mode (reader version 1.05 and higher) prior to any sample analysis. If out of range, manufacturer should be contacted for corrective action, 800-343-2170.
- __________
Do not proceed if out of range.
- __________
Reagent stability
- __________
Label Test Strips, one for each test sample and each
control
- __________
Place strip into appropriate incubator, avoid crushing sample compartment. For multiple samples,complete steps b-f for each sample/control, before starting test of next sample.
- __________
While holding strip flat, peel (to 'peel to here'
line) back plastic to expose sample pad compartment
- __________
Mix milk sample or control 25 times in seven seconds with a 1 ft movement or vortex control for 10 seconds when appropriate
- __________
Slowly pipette 300 µL sample to either side of the sample pad compartment(if applicable, use etched line on incubator as location guide). Avoid contact of sponge with pipet tip.
- __________
Draw up sample, avoiding foam and bubbles.(with 300ul calibrated pipet only pre-wet tip by
expelling sample and drawing up again)
- __________
Or for Screening only use poly pipet
- __________
While holding poly pipet almost vertically with bulb side pointing down, depress topbulb and insert into sample (below foam).
- __________
Release bulb and sample should completelyfill pipet shaft and overflow into bottom ½ of lower bulb
- __________
Pipet sample in pipet shaft by depressing top bulb only into test strip while leaving excess sample in overflow bulb
- __________
Re-seal plastic firmly around sample pad compartment
- __________
Repeat Step 7b-f for each sample and/or control test
strip in incubator
- __________
Close and latch incubator cover. If no automatic timer in incubator, set timer for 8 minutes, do not exceed 10 minutes
- __________
At end of incubation, visually inspect C (Control) line, an absent C line or a partial C line or an indistinct C line indicates an invalid test and the
sample/control must be re–tested
- __________
Insert only valid test(s) in ROSA reader (set to appropriate channel.
- __________
SLBL slow blink for SL Beta–lactam Test
- __________
SLBL rapid Blink for SL–6 Beta–lactam Test
- __________
Press ENTER, reading and interpretation appear in 5 seconds, read strips within 10 minutes of completion of incubation
- __________
Interpretation with ROSA Reader
- __________
If there is a negative or zero reading on the reader, sample is Negative (NF)
- __________
If there is a positive reading on the reader, sample is an Initial Positive
- __________
Verification of Initial Positive Samples (see App. N GR item 11); Confirmation of PresumptivePositive Samples (see App. N GR item 12); and Producer Traceback (see App. N GR item 13)
- __________
Reporting (see App. N GR item 14) .
- __________ Reader tapes or computer print outs maintained