FDA Home Page |
CFSAN Home | Search/Subject Index |
Q & A | Help
June 2, 2005
Milk Safety References
National Conference on Interstate Milk Shipments (NCIMS)
PHOSPHATASE TEST - FLUOROPHOS ALP TEST SYSTEM
[Unless otherwise stated all tolerances are ±5%]
Sample
- __________ Laboratory Requirements (see CP, item 33 & 34)
APPARATUS
- __________ See CP, items 1-32 (as necessary)
- __________ Cuvette Heating Block
- __________
Thermostatically-controlled at 38±1C
- __________
NIST-traceable thermometer, calibration checked as
specified in CP item 3
- __________
Temperature checked and recorded daily
- __________ Pipettors, fixed volume or electronic
- __________
75 µL pipettor
- __________
25 µL pipettor, for use with high-turbidity or high fat products (if needed)
- __________
Calibrated as specified in CP item 6e; records
maintained
- __________ Reagent Dispenser
- __________
Fixed volume 2.0 mL, calibrated and checked
- __________
Optionally, use 2.0 mL fixed volume or electronic
pipettor to dispense reagent
- __________
Calibrated as specified in CP item 6e; records
maintained
- __________ Cuvettes
- __________
Disposable glass 12 x 75 mm, dirt and scratch free
- __________ Fluorometer
- __________ Air fan in the rear unobstructed
- __________ Vents in the bottom base plate are unobstructed
- __________ Sufficient paper is on the roll in the printer
- __________ User's manual available
- __________Water Bath, 34±1C and 63±1C, circulating (Confirmation
procedures)
REAGENTS
- __________ Reagents, Handling and Storage
- __________
Test Reagent Set
- __________
Fluorophos substrate and Substrate buffer
- __________
Lot # _______ Rcd date _______ Exp date _______
- __________
Calibrator Set
- __________
Calibrators A, B and C
- __________
Lot # _______ Rcd date _______ Exp date _______
- __________
PhosphaCheck Pasteurization Controls Set
- __________
Positive and negative control
- __________
Lot # _______ Rcd date _______ Exp date _______
- __________
Daily Instrument Control
- __________
Lot # _______ Rcd date _______ Exp date _______
- __________
Reagents stored at 0-6C
- __________
Bottles labeled with receive and open dates
REAGENT PREPARATION
- __________ Working substrate
- __________ Prepare reagents as per manufacturer instructions,
mix by inversion until fully dissolved
- __________ Date (mixture stable 60 days at 0-6C)
- __________
Bottle labeled with date prepared
- __________
Preparation date _______
- __________ Place cleaned, 2 mL reagent dispenser (item 5) on
prepared reagent bottle, or cap if using 2 mL pipettor
CALIBRATION
(Required at installation and after any instrument adjustments)
- __________ Procedure
- __________
Perform instrument and reagent checks (item 14) prior
to proceeding
- __________
1. Readings from item 14 are within specification,
proceed with calibration
- __________
If readings not within specification, do not proceed
with calibration, make appropriate adjustments or
seek technical assistance and re-check
- __________
Record all values (initial and re-checks) on tape
and in QC record
- __________
Check calibration ratio of Calibrators A, B and C,
record Calibrator Lot # on strip and in QC record
- __________
1. Label two (2) tubes each for Calibrators A, B and C
- __________
Add 2 mL of each calibrator to the appropriately
labeled tubes
- __________
Heat tubes for 20 minutes to 38±1C
- __________
Find an empty channel
- __________
Press the "Calib" key _
- __________
Locate the channel to be used, using the "<" and ">" keys
- __________
Press the "Enter" key to select the channel
- __________
Place a tube of warmed Calibrator A (with no milk
added) into the cuvette chamber, close the door
and press the "Start" key
- __________
Continue as prompted until all six (6) tubes have been run
- __________
Calibration ratio should be 151±7 (when A/D mode check for Calibrator C/Daily Instrument Control is 602±6)
- __________
If ratio within specification continue, if not make adjustment and re-check calibration ratio
- __________
Invert retail containers 25 times, each inversion a
full cycle down and up
- __________
Remove test portions (avoiding foam) within 3 min
of agitation
- __________
To each calibrator, add 75 µL (or 25 µL) of the well-mixed product being tested
- __________
Prior to pipetting sample, draw up MS water
once and expel to waste
- __________
Dry exterior of piston and tip
- __________
Place tip of pipettor into sample (no more than
1 cm) and draw up and expel several times
- __________
Draw sample into pipettor, touch off to side
of container
- __________
Holding pipettor at 90° to lab bench and with tip
at eye level, dry exterior of tip by quickly wiping
from the pipettor over the tip
- __________
Carefully inspect the pipettor tip to insure
sample volume is flush with the tip
- __________
If concave, re-sample
- __________
If convex, re-wipe as above to achieve a flush
sample volume
- __________
Dispel 75 µL (or 25 µL) of sample 1 cm below the surface of the calibrator (do not dispense down side of cuvette)
- __________
With tip still below surface depress plunger three times into calibrator to completely expel sample
- __________
With plunger still completely depressed, remove from tube
- __________
Add products to calibrators one tube at a time
just prior to being tested
- __________
Mix by vortexing
- __________
Place cuvette in Fluorometer, close cuvette door
- __________
Press the "CALIB" key on the Fluorometer keypad and follow the prompts on
the display panel
- __________
After each reading, remove cuvette and close door
immediately
- __________
Record lot #'s of the calibrators used on the tape
printout and in the QC record book
- __________
Instrument calibrated for each product type to be
tested, some products with similar fat content may
share same channel
- __________
Re-calibration required if:
- __________
Controls out of limits
- __________
Adjustments made to bring A-D mode checks (item 14)
into specification
- __________
Any significant instrument service if performed,
ex. lamp or filter replaced
- __________
Instrument checks and calibrations within specification
CONTROLS
- __________ Negative Control
- __________ Use PhosphaCheck negative control from set in item 9c
- __________ Or, optionally heat a sample of product to 95±1C for 1 min with stirring (temperature control [TC] used)
- __________
Cool rapidly to 0-4.4C in an ice bath
- __________
If desired, distribute 1 mL quantities in small
tubes, seal and freeze in a non-frost-free freezer,
or place in a styrofoam container and place in the
center of a frost-free freezer for no more than
2 months
- __________ Add 2.0 mL of working substrate (Reagent C) to cuvettes
and heat to 38±1C for 20 min (use within 4 hours)
- __________ Add 75 µL of well mixed control to cuvette and immediately vortex
- __________ Once control is added to reagent test within 20 sec
- __________ Place the cuvette in the Fluorometer,
close the
cuvette door and press the "TEST" key on the keypad
- __________ Value less than (<) 20 mU/L
- __________ Positive Control
- __________
Use PhosphaCheck positive control from set in item 9c
- __________
Or, optionally to a portion of negative control, add
exactly 0.1 mL of mixed-herd raw milk and bring up to
exactly 100 mL with additional negative control
(as in item 12b)
- __________
If desired, distribute 1 mL quantities in small
tubes, seal and freeze in a non-frost-free freezer,
or place in a styrofoam container and place in the
center of a frost-free freezer for no more than
2 months
- __________
Test as in items 12 c-f
- __________
Value between 500±150 mU/L
INSTRUMENT AND REAGENT CHECKS
- __________ Optionally use manufacturer's reagent kits and instructions
specific for each instrument
- __________
Check calibration if readings are suspect (positive
control value out of limits)
- __________
Press set-up button
- __________
Press left or right arrow key to get A/D option
- __________
Select A/D Mode
- __________
Zero Check
- __________
With no tube in the instrument, press "Start" key and take a reading
- __________
A/D value
____________
- __________
The reading must not exceed 314. If the reading exceeds 314, an instrument problem is indicated, call for technical assistance
- __________
Record value on printout and QC record
- __________
Calibrator C/Daily Instrument Control Check
- __________
Dispense 2.0 mL of Calibrator C (item 9b) or Daily
Instrument Control (item 9d) into a 12x75 mm
cuvette and allow to warm to 38±1C for 20 minutes
- __________
Place the cuvette with the warmed Calibrator C or
Daily Instrument Control into the sample chamber
and press "Start" key
- __________
The A/D value should be 602±15 (maximum allowable drift.
- __________
A/D value ____________
- __________
Record Lot # and value on printout and QC record
- __________
If the value does not fall within the acceptable
range, then perform the following procedure (refer
to manual, or contact manufacturer if unsure)
- __________
With Calibrator C or Daily Instrument Control in
the sample chamber, adjust the R15 resistor until
the A/D value reads 602±2 .
- __________
Allow the instrument to equilibrate for 15 minutes, the A/D value should still read 602±2
- __________
Record Lot # and value on strip and in QC record
- __________
If the value does not fall within and stabilize at
602±2 seek technical assistance
- __________
For older units requiring the instrument cover to be removed or if unsure seek technical assistance
- __________
Reconstituted Substrate/Buffer stability check
- __________
Dispense 2.0 mL of working substrate (item 10)
into a 12x75 mm cuvette and allow to warm to 38±1C for 20 minutes
- __________
Place the cuvette with the warmed working substrate
into the sample chamber and press "Start" key
- __________
The A/D value should be < 1,200
- __________
A/D value ____________
- __________
Record Lot # and value on printout and QC record
- __________
Reconstituted Substrate/Buffer contamination check
- __________
Dispense 2.0 mL of working substrate (item 10)
into a 12x75 mm cuvette and allow to warm to 38±1C for 20 minutes
- __________
Place the cuvette with the warmed working substrate
into the sample chamber
- __________
Initiate an ALP sample reading of the working
substrate on an unused channel
- __________
The ALP value should be < 10 mU/L
- __________
ALP value __________
- __________
Record Lot # on printout and QC record
- __________
If the working substrate value does not fall within
the acceptable range, do not use working substrate,
re-check to verify, reconstitute a new set of
reagents or seek technical assistance before testing
samples
TEST PROCEDURE
- __________ Test Procedure
- __________
Perform all instrument and reagent checks (item 14),
negative control test (item 12) and positive control
test (item 13) prior to running analysis
- __________
Using reagent dispenser, fixed volume or electronic
pipettor, dispense 2.0 mL of working substrate into
labeled 12 x 75 mm glass cuvettes
- __________
Prime reagent dispenser (item 5) 3x prior to
dispensing volumes to cuvettes to remove any
bubbles from dispenser tubing
- __________
Warm substrate to 38±1C in the heating block for
20 min (use within 4 hours)
- __________
Sample agitation
- __________
Invert filled retail containers 25 times,
each inversion a full cycle down and up
- __________
Remove test portions within 3 min of agitation
- __________
Press the "Test" key on the keypad
- __________
Select the product type channel and enter identifi-
cation number
- __________
Dispense 75 µL (or 25 µL) of the well-mixed sample into
the warmed substrate and immediately mix by vortexing
- __________
Prior to pipetting sample, draw up MS water
once and expel to waste
- __________
Dry exterior of piston
- __________
Place tip of pipettor in sample (no more than
1 cm) and draw up and expel several times
- __________
Draw sample into pipettor, touch off to side
of container
- __________
Holding pipettor at 90° to lab bench and with tip
at eye level, dry exterior of tip by quickly wiping
from the pipettor over the tip
- __________
Carefully inspect the pipettor tip to insure
sample volume is flush with the tip
- __________
If concave, re-sample
- __________
If convex, re-wipe as above to achieve a flush
sample volume
- __________
Dispel 75 µL (or 25 µL) of sample 1 cm below the surface of the substrate (do not dispense down side of cuvette)
- __________
With tip still below surface depress plunger three times into substrate to completely expel sample
- __________
With plunger still completely depressed, remove from tube
- __________
Run test within 20 sec of adding sample to reagent
- __________
Place the cuvette in the Fluorometer, close the
cuvette door, and press the "START" key on the keypad
- __________
Results will display in 3 min., save tape print-
out of results in record book
- __________
If a 25 µL sample volume was used multiply the displayed value by 3
- __________
Record adjusted value on printout
- __________
Values of ³ 350 mU/L or more of ALP activity are considered to contain approximately 0.1% (v/v) raw milk and must be confirmed
CONFIRMATION
- __________ Negative Control
- __________
Prepare separate control for each product
- __________
Prepare by heating sample for at least 1 min
after thermometer registers 95±1C, stirring or mixing as necessary (TC used)__________
- Cool rapidly to 0-4.4C in an ice bath
- __________
This control must be less than 20 mU/L when tested
- __________ Positive Control (See item 13)
- __________
Must be prepared from suspect product
- __________ Microbial Phosphatase
- __________
To determine presence of microbial phosphatase,
heat 1.0 mL of suspect milk at 63±1C for 30 min,
stirring or mixing every 10 min (if fat content is
> 10%, heat at 66±1C) [TC used]
- __________
Cool rapidly to 0-4.4C in an ice bath
- __________
Test heated portion, unheated portion, and positive
and negative controls
- __________
Interpretation
- __________
If heated and unheated portions have equal activity
(within ±5%), the sample is regarded negative for
residual phosphatase, the activity originally
measured is microbial
- __________
If the heated portion has significantly reduced
(>5%) or no activity, the sample contains milk phosphatase activity, either
residual or reactivated
- Reactivated Phosphatase
- __________
Magnesium acetate solution
- __________
Dissolve 35.4g of Mg(C2H302)2·4H20
in 25 mL
MS water warming slightly to aid solution.
Pour solution into 100 mL volumetric flask,
rinse original container several times and
add rinses to flask. After cooling, make up
to 100 mL (stable for 1 year at 0-4.4C)
- __________
Procedure
- __________
Place 10 mL of each milk or milk product sample
to be tested in a boiling water bath and hold
1 min after temperature sample has reached
95±1C (TC used)
- __________
Cool samples rapidly to 0-4.4C in an ice bath
- __________
Place a 5 mL aliquot of sample (unheated) to be
tested in a screw-cap test tube and add 0.1 mL
MS water ("Blank" sample)
- __________
To a second 5 mL aliquot (unheated) in an
identical tube, add 0.l mL Mg acetate solution
("Test" sample)
- __________
Cap tubes and incubate both aliquots for 1 hr
at 34±1C
- __________
Remove samples from water bath and cool rapidly
to 0-4.4C in an ice bath
- __________
Dilute 1 mL of sample containing magnesium (Test)
with 5 mL (1:6 dilution) of corresponding boiled
milk or milk product control (items 21b 1 & 2 above)
- __________
Test undiluted sample containing no magnesium (Blank)
and diluted sample containing magnesium (Test) for
phosphatase activity (as described in item 16)
- Interpretation
- __________
If the diluted aliquot containing magnesium (Test)
has equal (±5%) or greater phosphatase activity
than the undiluted aliquot containing no magnesium
(Blank), the sample is regarded negative for residual
phosphatase, and the phosphatase originally measured
is of reactivated origin
Dil. w/Mg (Test) ³ Undil. (Blank) = Reactivated
- __________
If the diluted aliquot (Test) contains less
activity (< 5%) than the undiluted aliquot (Blank), the sample is considered
positive for residual phosphatase
Dil. w/Mg (Test) < Undil. (Blank) = Residual
- __________
A false-positive for residual phosphatase may
also be obtained if a reactivatable sample has
been allowed to stand at elevated temperatures
(20C) for periods of 1 hr or more before
testing (SPC <20,000/mL)
REPORT
- __________ Confirmatory Interpretation
- __________
Report as positive for residual phosphatase if
microbial phosphatase, and reactivatable phosphatase
are not present
- __________
Run standard/control 3x
- __________
Report as Not Found for residual phosphatase if:
- __________
If microbial phosphatase present
- __________
Or, if reactivatable phosphatase present
- __________
Or, if there is documentation that the product
was treated such that reactivatable phosphatase
may be present
- __________
Count obtained x 1000 is the cell count/mL milk
- __________
In reporting electronic somatic cell counts (ESCC/mL),
record only first two left hand digits, raising
second digit to next higher number when third digit
is 6 or more
- __________
Report the two left hand digits (rounded)
- __________ If the third digit is 5 the second digit is rounded
by the following rule
- __________ When the second digit is odd round up, raise the
second digit by 1 (odd up, 235 to 240)
- __________ When the second digit is even round down, delete
the 5 and report the second digit as is (even
down, 225 to 220)
- __________
If count on instrument is < 100 report as < 100,000 ESCC/mL