ELECTRONIC SOMATIC CELL COUNT
Foss 90
[Unless otherwise stated all tolerances are ±5%]
- __________ Laboratory Requirements (see CP, item 33 & 34)
- __________
Un-preserved samples may be run from 24 to 72 hours
after initial collection
- __________
Samples may be run from 8 hours up to 7 days after
initial collection if preserved with 0.02% 2-bromo-
2-nitropropane-1,3-diol (BronopolTM)or 0.05% potassium
dichromate (K2Cr2O7)
- __________
Comparative test with DMSCC
- __________
Performed by each analyst performing ESCC test
- __________
Test 4 samples (100K-200K, 300K-500K, 600K-800K
and 900K-1.2M) in triplicate for both DMSCC
(three separate smears each) and ESCC (three
separate sub-samples each, do not read same
sample three times)
- __________
Results must be shown to be acceptable prior
to official testing by analyst performing
comparison, i.e. analyst is not certified until
found acceptable. (co-requisite for certification)
- __________
Copy of comparison and results in QC record (or
easily accessible file in laboratory)
- __________
Analysts certified in DMSCC
APPARATUS
- __________ See Cultural Procedures, items 1–5
- __________ Fossmatic 90 Electronic Somatic Cell Counter
- __________ Pipettor, fixed volume or electronic (______________________)
- __________
Calibrated to deliver 500 µL milk (see CP item 6e)
- __________
Records maintained
- __________ Pipettor Tips
- __________
Disposable, replace for each sample
- __________
Reusable
- __________ Rinse in 40C deionized (DI) or MS water
- __________ Rinse in sample more than 1 time
- __________ Do not use for more than 25 samples
- __________ Water Bath
- __________
Circulating and thermostatically controlled to 37–42C
REAGENTS
- __________ Stock Dye Solution, 0.1% Ethidium Bromide (caution TOXIC,
use gloves when handling and do not breath dust)
- __________ Dissolve 1.0g ethidium bromide (C21H20BrN3) in 1 liter
DI or MS water by heating to 40-60C
- __________ Store in light-proof, air-tight bottle no more than
60 days
- __________ Date prep. ________ Exp. Date ________
- __________All stock dye/buffer and detergent solutions labeled with
date prepared and expiration date
- __________ Dissolve 10 mL Triton X-100 in 1 liter DI or MS water by
heating to 60C.
- __________ Store in air-tight container no more than 25 days
- __________ Date prep. ________ Exp. Date ________
- __________ Dye/Buffer Solution
- __________
Dissolve 51.0g KH phthalate and 13.75g KOH in 10 L
DI or MS water by heating to 40–60C
- __________
Add 150 mL 1% Triton X-100 (item 8), store less than
7 days in airtight container
- __________
Date prep. ________ Exp. Date ________
- __________ Ammonium Hydroxide (NH4OH) Solution,
Reagent Grade, 25%
- __________ All stock dye/buffer/rinsing solutions labeled with date
prepared and expiration date
WORKING SOLUTIONS
- __________ Working Dye Solution/Zero Control (used within 7 days)
- __________ Dilute 26 mL stock dye solution (item 7a) to
2.5 liter with stock buffer solution (item 9b2)
- __________ Date prep. ________ Exp. Date ________
- __________ Cell Counter
- __________
Add 10 mL stock rinsing solution (item 8) to 25 mL
of 25% NH4OH and dilute to 10 liters with DI or MS water
- __________
Date prep. ________ Exp. Date ________
- __________ Optionally use manufacturer's reagent kits and instructions
specific for each instrument
START UP
- __________ All working dye and rinsing solutions labeled with date
prepared and expiration date
- __________ Cell Counter
- __________
Assure adequate volume of working solutions, not
used beyond expiration date(s)
- __________
Turn on power and cycle at least six times__________
- Blind count
- __________
Vacuum pressure setting minimum of -40 KPa
- __________
Dispenser filling time 4–5 seconds
- __________
Intake filling time 3–4 seconds
- __________
IF ANY ABOVE PARAMETERS ARE WRONG, CORRECT BEFORE
PROCEEDING
- __________
Records maintained on all parameters
- __________ Milk Standards
- __________
Commercially prepared: ____________________
Lot# ______ Date Rcd. ______
- __________ Four samples in ranges 100K–200K, 300K–500K, 600K–
800K and 900K–1.2M
- __________ Do DMSCC in triplicate on each standard in set and
average counts, records maintained
- __________ DMSCC check performed in rotation by all certified
analysts
- __________ Standards used within one week
- __________
Certified provider: ___________________
- __________ Four standards in ranges 100K–200K, 300K-500K, 600K–
800K and 900K–1.2M
- __________ Maintain copies of all provided DMSCC values
- __________ Measure and maintain records of temperature (0–7.2C)
of standards as received
- __________ Maintain copies of all correspondence regarding
problems
- __________ Standards used by manufacturer's expiration date
__________
- Laboratory prepared (weekly)
- __________ Prepare from raw milk > 18 hours
old preserved with
0.05% potassium dichromate (K2Cr2O7)
- __________ Or, preserved with 0.02% 2–brono-2–nitropropane-
1,3–diol (Bronopol™)
- __________ Standards cannot be preserved with formalin
- __________ Prepare 4 standards in ranges 100K–200K, 300K–500K,
600K–800K and 900K–1.2M, used within one week
Date prep. ________ Exp. Date ________
- __________ Do DMSCC in triplicate on each standard prepared
and average counts, records maintained
- __________ DMSCC check performed in rotation by all certified
analysts
- __________
Hourly Control Sample (instrument drift check)
- __________ Use one of the standards (items 17a or b) in the
500–800K range, run in triplicate and determine
average
- __________ Optionally, prepare sufficient control/sample 500–800K range, run in triplicate and determine average
PROCEDURE
- __________ Testing Standards (each time instrument used)
- __________
Heat standards to 37–42C (using a temperature control)
and read within 30 minutes of reaching temperature,
used once and then discarded, i.e., do not re-use
- __________
Invert 10 times, pipet 500 µL into intake chamber within 3 min
- __________
Run standards in triplicate and average the counts
for each level, records maintained
- __________
Each standard's average must be within 10% of the
DMSCC (item 17) for that level, except within
15% for 100–200K standard, records maintained
- __________
Repeatability – a standard in the 300K to 800K range
must have a coefficient of variation (CV) of 5% or
less on 10 replicates (Refer to Operating Manual),
records maintained
- __________
THESE PARAMETERS MUST BE ACHIEVED BEFORE PROCEEDING__________
- Testing Samples
- __________
Heat samples to 37–42C (using a temperature control) and
read within 30 minutes of reaching temperature, samples
must not be re-used and must be discarded after use
- __________
Invert 10 times, pipet 500 µL into intake chamber within 3 min
- __________Record number of cells counted for each sample
- __________ With continuous operation:
- __________
Run a standard or optionally a control/sample (item 17d)
in the 500K to 800K range hourly, must be within 5% of
the original established instrument average value
(optionally, within 10% of original DMSCC average)
- __________
Run standard/control 3x
- __________
Run zero control (in item 12)
- __________ Maintain records
- __________ Routine maintenance
- __________
Perform as described in operating manual
- __________
Maintain records
REPORTS
- __________ Computing and Reporting Counts
- __________
Count obtained x 1000 is the cell count/mL milk
- __________
In reporting electronic somatic cell counts (ESCC/mL),
record only first two left hand digits, raising
second digit to next higher number when third digit
is 6 or more
- __________
Report the two left hand digits (rounded)
- __________ If the third digit is 5 the second digit is rounded
by the following rule
- __________ When the second digit is odd round up, raise the
second digit by 1 (odd up, 235 to 240)
- __________ When the second digit is even round down, delete
the 5 and report the second digit as is (even
down, 225 to 220)
- __________
If count on instrument is < 100 report as < 100,000 ESCC/mL