ELECTRONIC SOMATIC CELL COUNT
Somascope MKII
[Unless otherwise stated all tolerances are ±5%]
- __________ Laboratory Requirements (see CP, item 33 & 34)
- __________
Un-preserved samples may be run up to 72 hours
after initial collection
- __________
Samples may be run up to 7 days after initial collection
if preserved with 0.02% 2-bromo-2-nitropropane-1,3-diol
(Bronopol™) or 0.05% potassium dichromate (K2Cr2O7)
- __________
Comparative test with DMSCC
- __________
Performed by each analyst performing ESCC test
- __________
Test 4 samples (100K-200K, 300K-500K, 600K-800K
and 900K-1.2M) in triplicate for both DMSCC
(three separate smears each) and ESCC (three
separate sub-samples each, do not read
same
sample three times)
- __________
Results must be shown to be acceptable prior
to official testing by analyst performing
comparison, i.e. analyst is not certified until
found acceptable. (co-requisite for certification)
- __________
Copy of comparison and results in QC record (or
easily accessible file in laboratory)
- __________
Analysts certified in DMSCC
APPARATUS
- __________ See Cultural Procedures, items 1-5
- __________ Automated Electronic Somatic Cell Counters
- __________
Somascope MKII manual _____
- __________
Somascope MKII automatic _____
- __________ Water Bath
- __________
Circulating and thermostatically controlled to 37-42C
REAGENTS
- __________ Staining Kits
Warning: Before using the chemicals for making up the
staining solution, read the Material Safety Data Sheet
(MSDS) provided in the appendix of the manual, wear safety
glasses and latex gloves when handling the chemicals
- __________
One liter concentrate kit, makes 5 L staining
solution when diluted mixed with deionized (DI) or
MS water
__________Date Rcd _____ Exp. Date _____
- __________
Store at 0-5C and protected from light
- __________
Twenty liter powder kit, consists of three different
components, mixed with 20 L DI or MS water
__________Date Rcd _____ Exp. Date _____
- __________
Staining Concentrate (powder)
- __________
Store powder 0 - 5C and protected from light
- __________
Staining Detergent (liquid)
- __________
Store at room-temperatu˘re protected from light
- __________
Staining Buffer (powder)
- __________
Store at room-temperature protected from light
WORKING SOLUTIONS
- __________ Working Staining Solutions
- __________
One liter Concentrate Kit
- __________ Add 4 L of DI or MS water
- __________ Mix on a magnetic stirrer at room temperature
to make working stain solution
- __________
Twenty liter Powder Kit
- __________ Dissolve the staining buffer (item 5b3) in
about 18 L of DI or MS water and stir until the
powder is fully dissolved
- __________ Add the Staining Detergent (item 5b2) to about
one liter of warm (35 - 45C) DI or MS water
and mix well (preferably with a magnetic
stirrer) to dissolve the detergent
- __________ The detergent must be well dissolved, no
powder residue visible (optical test)
- __________ Add the detergent solution (item 6b2) to the
18 L of staining buffer (item 6b1) and mix
- __________ Dissolve the Staining Concentrate (item 5.b.1.) in
3 mL of 35-45C DI or MS water
- __________ Mix until the powder is dissolved (optical test)
- __________ Keep the concentrate (powder and solution)
protected from strong light during preparation
- __________ Add the dissolved concentrate to the approximately
19 L of buffer and add DI or MS water until the
total volume is 20 L
- __________
Store the working staining solutions up to 2 months
at 0 - 5C protected from light
__________
Date prep. ________ Exp. Date ________
- __________
Use the staining solution at room-temperature
__________
Date filled _____
- __________ The contents of the 5 liter staining container can
be left at room temperature
- __________ The contents must be used within 7 days
- __________ Clean container once a month as per manufacture
instructions
- __________ Working Rinse Solution
- __________ Detergent Container
- __________ Alkaline detergent; DECON 90, Contrad 70 or RBS 50
- __________ Fill the black detergent reservoir with about
50 mL of undiluted detergent in the Sample
Preparation Unit
- __________ Check reservoir visually daily for volume and
add more detergent as necessary
- __________ Water Container
- __________ Measure 5 mL of Triton X-100 and place in 100 mL
of DI or MS water
- Mix the above solution until the Triton X-100 is completely dissolved
- __________ Dilute the 100 mL mixture above to 5 liters with
room temperature DI or MS water
- __________ Mix well
- __________ Pour the above into the water container provided
with the instrument
START UP
- __________ Somatic Cell Counter
- __________ Check to see if the amount of staining solution
(about 1.6 mL per sample), detergent and rinse
solution (about 1 mL per sample) is of sufficient
volume for the number of samples to be run
- __________ Solutions not be used beyond expiration date
- __________ Turn on the HBO lamp, Sample Preparation Unit and
wait 15 minutes
- __________ Turn on the internal computer of the Somascope MKII
- __________ Put the peristaltic pump in operation position
- __________ Perform a clean sequence
- __________ Run at least 5 times with rinse solution (item 7b),
last reading must be below 5
- __________ IF ANY PARAMETERS ARE OUT OF TOLERANCE, CORRECT
BEFORE PROCEEDING
- __________ Records maintained on all parameters each time
instrument is used
- __________ Milk Standards
- __________
Commercially prepared: ____________________
__________ Lot# ______ Date Date Rcd. ______
- __________ Four standards in ranges 100K-200K, 300K-500K,
600K-800K and 900K-1.2M
- __________ Do DMSCC in triplicate on each standard in set and
average counts, records maintained
- __________ DMSCC check performed in rotation by all certified
analysts
- __________ Standards used within one week
- __________
Certified provider: ____________________
__________ Lot# ______ Exp.Date Rcd. ______
- __________ Four standards in ranges 100K-200K, 300K-500K,
600K-800K and 900K-1.2M
- __________ Maintain copies of all provided DMSCC values
- __________ Measure and maintain records of temperature
(0-7.2C) of standards as received
- __________ Maintain copies of all correspondence regarding
problems
- __________ Standards used by manufacturer's expiration date
- __________
Laboratory prepared (weekly)
- __________ Prepare from raw milk > 18 hours
old preserved with 0.05% potassium dichromate (K2Cr2O7)
- __________ Or, preserved with 0.02% 2-brono-2-nitropropane-
1,3-diol (Bronopol™)
- __________ Standards cannot be preserved with formalin
- __________ Prepare 4 standards in ranges 100K 200K, 300K 500K,
600K-800K and 900K 1.2M, use within one week
Date prep. ________ Exp. Date ________
- __________ Do DMSCC in triplicate on each standard prepared
and average counts, records maintained
- __________ DMSCC check performed in rotation by all certified
analysts
- __________
Hourly Control Sample (instrument drift check)
- __________ Use one of the standards (items 11a or b) in
the 500-800K range, run in triplicate and
determine average
- __________ Optionally, prepare sufficient control/sample
500-800K range, run in triplicate and determine
average
PROCEDURE
- __________ Testing Standards (each time instrument used)
- __________
Heat standards to 37-42C (using a temperature control)
and read within 30 minutes of reaching temperature,
used once and then discarded, i.e., do not re-use
- __________
Mix by inverting at least 2x, test standards
within 3 minutes
- __________
Run the standards in triplicate and average the
counts for each level, records maintained
- __________
Each standard's average must be within 10% of the
DMSCC (item 11) for that level, except within 15%
for 100-200K standard, records maintained
- __________
Repeatability - a standard in the 300K to 800K range
must have a coefficient of variation (CV) of 5% or
less on 10 replicates, records maintained
- __________
THESE PARAMETERS MUST BE ACHIEVED BEFORE PROCEEDING
- __________ Testing samples
- __________ Heat samples to 37-42C (using a temperature control)
and read within 30 minutes of reaching temperature;
samples must not be re-used and must be discarded
after use
- __________ Mix by inverting at least 2x, test samples within
3 minutes
- __________ Samples must be tested within 10 minutes after
being removed from waterbath
- __________ Samples must not be reused and must be discarded
after use
- __________ Record number of cells counted for each sample
- __________ With continuous operation:
- __________ Run a standard or optionally a control/sample
(item 9d) in the 500K to 800K range hourly, must
be within 5% of the original established instrument
average value (optionally, within 10% of original
DMSCC average)
- __________ Run control 3x
- __________ Run zero control (item 8g) hourly
- __________ Maintain records
- __________ Shut down procedure
- __________
Perform a clean cycle twice
- __________
Put instrument pipette in beaker of rinse solution
- __________
Release peristaltic pump
- __________
Switch off PC, SPU, internal computer of Somascope
MKII, HBO-lamp in this order
- __________ Routine maintenance
- __________
Perform as described in operating manual
- __________
Maintain records
REPORTS
- __________ Computing and Reporting of Counts
- __________
Count obtained x 1000 is the cell
count/mL milk
- __________
In reporting electronic somatic cell counts
(ESCC/mL), record only first two left hand
digits, raising second digit to next higher
number when third digit is six or more__________
- Report the two left hand digits (rounded)
- __________ If the third digit is 5 the second digit is
rounded by the following rule
- __________ When second digit is odd round up, raising
the second digit by 1 (odd up, 235 to 240)
- __________ When second digit is even round down,
delete the 5 and report the second digit
as is (even down, 225 to 220)
- __________
If count on instrument is < 100 report count as < 100,000 ESCC/mL