ELECTRONIC SOMATIC CELL COUNT
Fossomatic 5000
[Unless otherwise stated all tolerances are ±5%]
Sample
- __________ Laboratory Requirements (see CP, item 33 & 34)
- __________
Un-preserved samples may be run up to 72 hours
after initial collection
- __________
Samples may be run up to 7 days after initial collection
if preserved with 0.02% 2-bromo-2-nitropropane-1,3-diol
(BronopolTM) or 0.05% potassium dichromate (K2Cr2O7)
- __________
Comparative test with DMSCC
- __________
Performed by each analyst performing ESCC test
- __________
Test 4 samples (100K-200K, 300K-500K, 600K-800K
and 900K-1.2M) in triplicate for both DMSCC
(three separate smears each) and ESCC (three
separate sub-samples each, do not read same
sample three times)
- __________
Results must be shown to be acceptable prior
to official testing by analyst performing
comparison, i.e. analyst is not certified until
found acceptable. (co-requisite for certification)
- __________
Copy of comparison and results in QC record (or
easily accessible file in laboratory)
- __________
Analysts certified in DMSCC
APPARATUS
- __________ See Cultural Procedures, items 1-5
- __________
Water Bath
- __________
Circulating and thermostatically controlled to 37-42C
REAGENTS
- __________
Preparation
- __________
Fixed volume or electronic, 100 µL
- __________
Stock Solution: Dissolve 500 mL of Clean 5000
in 4.5 L of deionized (DI) or MS water, heat to
about 60C, store in airtight, light proof container
in a cool location and use within 16 weeks
__________
Date Prep. ________ Exp. Date ________
- __________
Buffer/diluent Solution: Mix 1 L of stock solution
with one bag (354 g) of Buffer 5000, add DI or MS
water to 10 L, heat to 40 - 60C to speed process,
store in diluent container next to instrument and
use within 3 weeks (enough for about 25,000 samples)
__________
Date Prep. ________ Exp. Date ________
- __________
Rinse/sheath Liquid: Mix 250 mL of stock solution
with DI or MS water to make 50 L, store and use
within 3 weeks (enough for about 5000 samples)
__________
Date Prep. ________ Exp. Date ________
- __________
Insert Dye 5000 bag according to manufacturer's
instructions
- __________
0.5 Liter Bags
- __________
Stock Solution: Dissolve 100 mL of Clean 5000 in
900 mL of DI or MS water (use syringe), heat to
about 60C, store in airtight, light proof container
in a cool location and use within 16 weeks
__________
Date Prep. ________ Exp. Date ________
- __________
Buffer/diluent Solution: Mix 0.5 L of stock solution
with one bag (171 g) of Buffer 5000, add DI or MS
water to 5 L, heat to 40 - 60C to speed process,
store in diluent container next to instrument and
use within 3 weeks(enough for about 10,000 samples)
__________
Date Prep. ________ Exp. Date ________
- __________
Rinse/sheath Liquid: Mix 100 mL of stock solution
with DI or MS water to 20 L, store and use within
3 weeks (enough for about 2000 samples)
__________
Date Prep. ________ Exp. Date ________
- __________
Insert Dye 5000 bag according to manufacturer's instructions
- __________
Stock Solutions for 5000 basic
- __________
Dye/buffer stock solution: Dissolve 3 ethidium
bromide tablets in 1 L of DI or MS water, stir to
completely dissolve tablets, store in dark cool
location, use within 16 weeks
__________
Date Prep. ________ Exp. Date ________
- __________
Clean 5000 stock solution: Dilute one bottle of
Clean 5000 in 4.5 L of DI or MS water, heat to
about 40 - 60C to speed process, store and use
within 16 weeks
__________
Date Prep. ________ Exp. Date ________
- __________
Working Solutions for 5000 basic
- __________
Buffer/diluent solution: Dissolve one bag of Buffer
5000 in about 8 L of DI or MS water in 10 L container,
add 1 L of Clean 5000 stock solution and fill to a
total of 10 L with DI or MS water, store and use
within 6 weeks
__________
Date Prep. ________ Exp. Date ________
- __________
Dye/buffer solution: Mix 1800 mL of Buffer/diluent
solution and 200 mL of Dye stock solution in the
reagent bottle and place in the instruments according
to instructions, use within 6 weeks
__________
Date Prep. ________ Exp. Date ________
- __________
Rinse/sheath liquid: Fill a 50 L container with
about 49 L of DI or MS water, add 250 mL of Clean
5000 stock solution and fill to 50 L with deionize
water, use within 3 weeks
__________
Date Prep. ________ Exp. Date ________
- __________
Storage Periods and Temperatures
- __________
Storage time for Buffer 5000 and Clean 500 is one year
from FE (Foss Electric, i.e., manufacture) date
__________
FE Date ________ Exp. Date ________
- __________
Store Dye 5000 at 2-8C for a maximum of 1 year from
FE date
__________
FE Date ________ Exp. Date ________
- __________
Rinse 5000, used to rinse dye interface when instrument
has not been used for more than 7 days, store for 1 year
from FE date, if seal is not broken, after seal is broken
store at 2-8C for no more than 6 months
__________
FE Date ________ Exp. Date ________
- __________
Other Liquids
- __________
Blind solution: Prepare a 1% (w/v) NaCl solution
- __________
Cleaning solution for milk waste tube: Prepare a 0.5%
S-470 solution (5 g per Liter), heat to 50 - 70C, used
to clean the milk waste tube at end of day, use within
1 week
Date prep. ________ Exp. Date
- __________
Cleaning solution for flow cell flush: Use 0.5% S-470
solution (see item b above)
- __________
All solutions labeled with date prepared and expiration date
START UP
- __________ Cell Counter
- __________
Turn power on and place instrument in standby mode
- __________
Check that enough rinse/sheath liquid, dye and buffer
solutions are available in the external containers
- __________
Solutions not used beyond expiration date(s)
- __________
Perform a blind check before starting measurement,
if mean count is < 3000 cells/mL and individual
measurements < 5000, within acceptable limits
- __________
IF ANY ABOVE PARAMETERS ARE WRONG, CORRECT BEFORE
PROCEEDING
- __________
Records maintained on all parameters each time
instrument is used
- __________ Milk Standards
- __________
Commercially prepared: ___________________
Lot# ______ Date Rcd. ______________
- __________
Stock Solution: Dissolve 500 mL of Clean 5000
in 4.5 L of deionized (DI) or MS water, heat to
about 60C, store in airtight, light proof container
in a cool location and use within 16 weeks
- __________
Do DMSCC in triplicate on each standard in set and
average counts, records maintained
- __________
DMSCC check performed in rotation by all certified analysts
- __________
Standards used within one week
- __________
Certified provider: ___________________
Lot# ______ Exp. Date ______ Date Rcd. ______
- __________
Four standards in ranges 100K-200K, 300K-500K, 600K-
800K and 900K-1.2M
- __________
Maintain copies of all provided DMSCC values
- __________
Measure and maintain records of temperature (0-7.2C)
of standards as received
- __________
Maintain copies of all correspondence regarding problems
- __________
Standards used by manufacturer's expiration date
- __________
Laboratory prepared (weekly)
- __________
Prepare from raw milk > 18 hours old preserved with
0.05% potassium dichromate (K2Cr2O7)
- __________
Or, preserved with 0.02% 2-bromo-2-nitropropane-
1,3-diol (Bronopol™)
- __________
Standards cannot be preserved with formalin
- __________
Prepare 4 standards in ranges 100K 200K, 300K 500K,
600K-800K and 900K 1.2M, used within one week
__________
Date prep. ________ Exp. Date _________
- __________
Do DMSCC in triplicate on each standard and average
counts, records maintained
- __________
DMSCC check performed in rotation by all certified
analysts
- __________
Hourly Control Sample (instrument drift check)
- __________
Use one of the standards (items 9a or b) in the 500-
800K range, run in triplicate and determine average
- __________
Optionally, prepare sufficient control/sample 500-
800K range, run in triplicate and determine average
PROCEDURE
- __________ Handling and storage
- __________
Heat standards to 37-42C (using a temperature control)
and read within 30 minutes of reaching temperature,
used once and then discarded, i.e. do not re-use
- __________
Mix by inverting at least 2x, place in rack and put
onto automatic track, run within 10 minutes
- __________
Run the standards three times and average the counts for
each level, records maintained
- __________
Each standard's average must be within 10% of the
DMSCC (item 9) for that level, except within 15%
for 100K-200K standard, records maintained
- __________
Repeatability - a standard in the 300K to 800K range
must have a coefficient of variation (CV) of 5% or
less on 10 replicates (Refer to Operating Manual),
records maintained
- __________
THESE PARAMETERS MUST BE ACHIEVED BEFORE PROCEEDING
CONTROLS
- __________ Testing Samples
- __________
Heat samples to 37-42C (using a temperature control) and
read within 30 minutes of reaching temperature; samples
must not be re-used and must be discarded after use
- __________
Mix by inverting at least 2x, place in rack and put
onto automatic track, run within 10 minutes of
reaching the testing temperature
- __________
With continuous operation:
- __________
Run a standard or optionally a control/sample (item 9d)
in the 500K to 800K range hourly, must be within 5% of
the original established instrument average value
(optionally, within 10% of original DMSCC average)
- __________
Run control 3x
- __________
Run zero control (item 8d)
- __________
Maintain records
- __________
Routine maintenance
- __________
Perform as described in operating manual
- __________
Maintain records
REPORTS
- __________
Computing and Reporting Counts
- __________
Count obtained x 1000 is the cell count/mL milk
- __________
In reporting electronic somatic cell counts (ESCC/mL),
record only first two left hand digits, raising
second digit to next higher number when third digit
is 6 or more
- __________
Report the two left hand digits (rounded)
- __________
If the third digit is 5 the second digit is rounded
by the following rule
- __________
When the second digit is odd round up, raise the
second digit by 1 (odd up, 235 to 240)
- __________
When the second digit is even round down, delete
the 5 and report the second digit as is (even
down, 225 to 220)
- __________
If count on instrument is < 100 report as < 100,000 ESCC/
mL