ELECTRONIC SOMATIC CELL COUNT
Foss 250/300/360/400
[Unless otherwise stated all tolerances are ±5%]
- __________ Laboratory Requirements (see CP, item 33 & 34)
- __________
Un-preserved samples may be run up to 72 hours
after initial collection
- __________
Samples may be run up to 7 days after initial collection
if preserved with 0.02% 2-bromo-2-nitropropane-1,3-diol
(BronopolTM) or 0.05% potassium dichromate (K2Cr2O7)
- __________
Comparative test with DMSCC
- __________
Performed by each analyst performing ESCC test
- __________
Test 4 samples (100K-200K, 300K-500K, 600K-800K
and 900K-1.2M) in triplicate for both DMSCC
(three separate smears each) and ESCC (three
separate sub-samples each, do not read
same
sample three times)
- __________
Results must be shown to be acceptable prior
to official testing by analyst performing
comparison, i.e. analyst is not certified until
found acceptable. (co-requisite for certification)
- __________
Copy of comparison and results in QC record (or
easily accessible file in laboratory)
- __________
Analysts certified in DMSCC
APPARATUS
- __________ See Cultural Procedures, items 1-5
- __________ Automated Electronic Somatic Cell Counters
- __________
Fossomatic 250
- __________
Fossomatic 300
- __________
Fossomatic 360
- __________
Fossomatic 400
- __________ Water Bath
- __________
Circulating and thermostatically controlled to 37-42C
REAGENTS
- __________ Stock Dye/Buffer Solution
- __________
Dissolve 2.5g (or number of tablets specified by manu-
facturer) ethidium bromide (C21H20BrN3) in 1 liter deionized
(DI) or MS water (caution TOXIC, use gloves when handling
and do not breath dust), heat to 40-60C and mix to dissolve
- __________
Add 400g tripotassium citrate monohydrate (C6H5O7K3·H2O),
14.5g citric acid monohydrate (C6H8O7·H2O),
and 4 liters DI or MS water, heat to 40-60C and mix to dissolve
- __________
Add dye and buffer solutions together and mix
- __________
Add 50 mL neutral detergent, Triton X-100 to mixture
and stir until dissolved
- __________
Store refrigerated (0-4.4C) in airtight, light-proof
container for no longer than 90 days
- __________
Date prep. ________ Exp. Date
- __________ Stock Detergent Solution
- __________
Dissolve 10 mL neutral detergent, Triton X-100 in
1 liter of DI or MS water and heat 40-60C to complete
solution
- __________
Store refrigerated (0-4.4C) in airtight, container
for no longer than 30 days
- __________
Date prep. ________ Exp. Date ________
- __________ Ammonium Hydroxide (NH 4 OH) Solution, Reagent Grade, 25%
- __________All stock dye/buffer and detergent solutions labeled with
date prepared and expiration date
WORKING SOLUTIONS
- __________ Dye/Buffer Solution
- __________
Dilute 1 L dye/buffer stock solution (item 5) with
9 L DI or MS water
- __________
Protect from light and use within 21 days
- __________
Date prep. ________ Exp. Date ________
PERFORMANCE TESTING
- __________ Rinsing Solution (use within 7 days)
- __________
Add 10 mL of stock neutral detergent stock solution
(item 6) and 25 mL of ammonium hydroxide solution
(item 7) and suspend to 10 L with DI or MS water
- __________
Date prep. ________ Exp. Date ________
- __________ Optionally, use manufacturer's reagent kits and instructions
specific for each instrument
- __________All working dye/buffer and rinsing solutions labeled with
date prepared and expiration date
START UP
- __________ Cell Counter
- __________
Check that the amount of dye/buffer solution (item 9)
and rinsing (cleaning) solution (item 10) in instrument
supply containers is of sufficient volume for the
number of samples to be run
- __________
Solutions not used beyond expiration date(s)
- __________
Turn on power and cycle at least six times
- __________
Perform a zero check before starting any measurements,
within acceptable limits, single counts up to 5 and
mean up to 3
- __________
IF ANY ABOVE PARAMETERS ARE WRONG, CORRECT BEFORE
PROCEEDING
- __________
Records maintained on all parameters each time
instrument is used
REPORTING AND RECORDS
- __________ Milk Standards
- __________
Commercially prepared: ___________________
Lot# ______ Date Rcd. ______
- __________
Four standards in ranges 100K-200K, 300K-500K, 600K-
800K and 900K-1.2M
- __________
Do DMSCC in triplicate on each standard in set and
average counts, records maintained
- __________
DMSCC check performed in rotation by all certified
analysts
- __________
Standards used within one week
- __________
Certified provider: ___________________
Lot# ______ Exp. Date ______ Date Rcd. ______
- __________
Four standards in ranges 100K-200K, 300K-500K, 600K-
800K and 900K-1.2M
- __________
Maintain copies of all provided DMSCC values
- __________
Measure and maintain records of temperature (0-7.2C)
of standards as received
- __________
Maintain copies of all correspondence regarding
problems
- __________
Standards used by manufacturer's expiration date
- __________
Laboratory prepared (weekly)
- __________
Prepare from raw milk > 18 hours old preserved with
0.05% potassium dichromate (K2Cr2O7)
- __________
Or, preserved with 0.02% 2-bromo-2-nitropropane-
1,3-diol (Bronopol™)
- __________
Standards cannot be preserved with formalin
- __________
Prepare 4 standards in ranges 100K 200K, 300K 500K,
600K-800K and 900K 1.2M, used within one week Date prep. ________ Exp. Date ________
- __________
Do DMSCC in triplicate on each standard and average
counts, records maintained
- __________
DMSCC check performed in rotation by all certified
analysts
- __________
Hourly Control Sample (instrument drift check)
- __________
Use one of the standards (items 14a or b) in the
500-800K range, run in triplicate and determine
average
- __________
Optionally, prepare sufficient control/sample 500-
800K range, run in triplicate and determine average
PROCEDURE
- __________ Testing Standards (each time instrument used)
- __________
Heat standards to 37-42C (using a temperature control)
and read within 30 minutes of reaching temperature,
used once and then discarded, i.e. do not re-use
- __________
Mix by inverting at least 2x, place in rack and put
onto automatic track, run within 10 minutes__________
- Run the standards six times and average the counts for each level, records maintained
- __________
Each standard's average must be within 10% of the
DMSCC (item 14) for that level, except within 15%
for 100K-200K standard, records maintained
- __________
Repeatability - a standard in the 300K to 800K range
must have a coefficient of variation (CV) of 5% or
less on 10 replicates (Refer to Operating Manual),
records maintained
- __________
THESE PARAMETERS MUST BE ACHIEVED BEFORE PROCEEDIN
- __________ Testing Samples
- __________
Heat samples to 37-42C (using a temperature control)
and read within 30 minutes of reaching temperature;
samples must not be re-used and must be discarded
after use
- __________
Mix by inverting at least 2x, place in rack and put
onto automatic track, run within 10 minutes of
reaching the testing temperature__________
- __________ With continuous operation:
- __________
Run a standard or optionally a control/sample (item 14d)
in the 500K to 800K range hourly, must be within 5% of
the original established instrument average value
(optionally, within 10% of original DMSCC average)
- __________
Run control 6x
- __________Run zero control (item 13d)
- __________
Maintain records
- __________ Routine maintenance
- __________
Perform as described in operating manual
- __________
Maintain records
REPORTS
- __________ Computing and Reporting Counts
- __________
Count obtained x 1000 is the cell count/mL milk
- __________
In reporting electronic somatic cell counts (ESCC/mL),
record only first two left hand digits, raising
second digit to next higher number when third digit
is 6 or more
- __________Report the two left hand digits (rounded)
- __________
If the third digit is 5 the second digit is rounded
by the following rule
- __________
When the second digit is odd round up, raise the
second digit by 1 (odd up, 235 to 240)
- __________
When the second digit is even round down, delete
the 5 and report the second digit as is (even
down, 225 to 220)
- __________
If count on instrument is < 100 report as < 100,000 ESCC/mL