APPENDIX N BULK MILK TANKER SCREENING TEST FORM
IDEXX - New SNAP® BETA LACTAM
For Raw Bovine Milk
GENERAL REQUIREMENTS
- __________
See Appendix N General Requirements form items 1-8 & 15
SAMPLES
- __________
See Appendix N General Requirements (GR) form item 9
APPARATUS & REAGENTS
- __________
Equipment
- __________
Heater block with SNAP inset thermostatically controlled at 45±5C
- __________
Temperature checked by placing standardized thermometer in tube containing liquid
(bulb submersed) in heating unit, records maintained
- __________
Or, use 6 inch partial immersion thermometer placed directly into small thermometer
well in middle of heating unit, records maintained
- __________
Single use 450 µL pipet with indicator line to measure amount of sample, supplied by manufacturer
(screening only)
- __________
Fixed volume pipettor to dispense 450 µL (see App. N GR item 7)
- __________
SNAP Kit
__________
Lot # _____ Exp Date _____
- __________
Sample tubes containing reagent pellet
- __________
Kits received refrigerated
- __________
Store kits at 0-7C
- __________
Timer
- __________
IDEXX Reader for SNAP devices, with printer or data download capability
- __________
Controls [FROZEN STANDARDS NOT ALLOWED]
- __________
Non frozen positive control only, 5.0 ppb Penicillin G
- __________
Store according to label instructions
__________
Mfg. _____ Lot # _____ Exp. Date _____
- __________
Re-hydrate as per manufacturer's instructions
- __________
Test for suitability each time prepared, must produce appropriate reaction, records maintained
- __________
Store control at 0 - 4.4C for no more than 2 days.
Do not freeze
- __________
Inhibitor Free Raw Milk (Negative Control)
- __________
Test for suitability each time prepared, must produce appropriate reaction, records maintained
TECHNIQUE
- __________
Daily Performance and Operation Check (see App. N GR item 10)
- __________
Read positive and negative check set devices.
Both devices must read within the limits on the check set devices
__________
Positive Range __________ Negative Range __________
__________
If check sets fail, call IDEXX before proceeding
- __________
Test Procedure
- __________
Set out required number of SNAP™ devices, sample tubes and pipets for the samples to be tested.
Discard unused, un-refrigerated devices at the end of the day
- __________
Pre-warm heater block(s) to 45±5C, hold at 45C for at least 5 minutes
- __________
Check initial pre-heating with a reference thermometer, records maintained
- __________
Continuous use block heaters, check temperature daily with reference thermometer, records maintained
- __________
Label each device and each sample tube
- __________
Place devices on incubator block(s)
- __________
Mix samples/controls by shaking 25 times in 7 sec through 1 ft arc, use within 3 minutes
- __________
Look for blue reagent pellet in bottom of tube, if not there tap to bring pellet down
- __________
Remove and discard sample tube caps
- __________
With pipets provided, draw up controls or samples
- __________
(Draw up, avoiding foam and bubbles, expel and draw up again) to the indicator lines and carefully
add all of the control or sample milk to the appropriately labeled tubes. (Screening only)
- __________
Or, using fixed volume pipettor (item 3c), draw up 450 µL of controls and samples (draw up,
avoiding foam and bubbles, expel and draw up again) and carefully add to the appropriately labeled tubes
- __________
Use clean pipet (or tip) for each control and sample
- __________
Agitate sample tube to dissolve reagent pellet
- __________
Incubate tube(s) in heater block next to device with the corresponding ID
- __________
Incubate tubes for 5 minutes (use timer) at 45±5C
- __________
After incubation, pour contents of tubes into sample well of device
- __________
Watch blue activation circle, as it begins to become white push the Activator firmly until it "snaps"
flush with the body of the SNAP™ device (device remains on heater block)
- __________
Incubate device for 4 minutes (use timer) at 45±5C
- __________
Read IMMEDIATELY with IDEXX Reader for SNAP devices
- __________
Interpretation
- __________
The control spot is on the top and the test spot on the bottom of the Results Window (Correct
orientation is with activator button to right and sample well to left)
- __________
Negative result:
- __________
If test spot is darker than or equal to the control spot, sample is Negative (NF)
- __________
Positive result:
- __________
If test spot is lighter than control spot, sample is Initial Positive
- __________
IDEXX Reader for SNAP devices automatically prints results as Positive (initial) or Negative (NF)
- __________
Verification of Initial Positive Samples (see App. N GR item 11);
Confirmation of Presumptive Positive Samples (see App. N GR item 12); and Producer Trace-Back (see App. N GR item 13)
- __________
Reporting (see App. N GR item 14)